Biosystems PhD Ultra Manual de Instruções

Consulte online ou descarregue Manual de Instruções para Detetores de fumo Biosystems PhD Ultra. biosystems PhD Ultra Instruction manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 0
1
Reference Manual
PhD Ultra
Multi Gas Detector
651 South Main Street
Middletown, CT 06457 USA
860 344-1079, 800 711-6776
FAX 860 344-1068
http://www.biosystems.com
Version 2.60 28JAN2008
Part Number 13-037
Vista de página 0
1 2 3 4 5 6 ... 59 60

Resumo do Conteúdo

Página 1 - Multi Gas Detector

1 Reference Manual PhD Ultra Multi Gas Detector 651 South Main Street Middletown, CT 06457 USA 860 344-1079, 800 711-6776 FAX 860 344-106

Página 2

10 Use of these procedures is reserved for authorized personnel. 1.5 Instrument identifiers The PhD Ultra includes two built-in instrument identifier

Página 3

11 1.6.3.1 Low battery alarm settings for NiCad battery packs The PhD Ultra includes low battery alarms that are activated whenever battery voltage

Página 4

12 1.6.4.5 Down-scale or negative reading alarms Significantly negative or “down-scale” readings cause the activation of PhD Ultra audible and visua

Página 5

13 pack is momentarily removed or replaced with another it will be necessary to manually turn the PhD Ultra off if the instrument is not going to be

Página 6

14 (3) LCD display: A "Supertwist" liquid crystal display (LCD) meter allows display of readings, messages, and other information. (4) Alar

Página 7 - Signal Words

15 Chapter 2 Basic operation 2.1 Operation overview 2.1.1 Turning the PhD Ultra on The large black push-button on the top of the PhD Ultra case is

Página 8 - Warnings

16 If the instrument is operated in the “Text Only” mode an “OK” message will be displayed as long as an alarm set point has not been exceeded. If t

Página 9 - 1.4 Calibration

17 LED alarm lights will flash and the audible alarm will sound during an alarm condition. PhD Ultra alarms are normally self-resetting. When readi

Página 10 - 1.6 Alarm logic

18 average concentration for the most recently completed 15 minutes of operation. Note: For the first 15 minutes after the PhD Ultra is initially

Página 11

19 2.3.1 NiCad battery pack When the PhD Ultra is operated in the diffusion mode, the rechargeable NiCad battery pack will provide up to 12 hours

Página 12 - 1.9.2 Batteries

2 THE PhD ULTRA PERSONAL PORTABLE GAS DETECTOR HAS BEEN DESIGNED FOR THE DETECTION OF OXYGEN DEFICIENCIES, FLAMMABLE GAS, AND TOXIC VAPOR ACCUMUL

Página 13 - Front Surfaces)

20 consists of two components a slip-in cradle and a 110 VAC “wall cube” type power source. European and Australian chargers may use a different wal

Página 14 - 1.11 PhD Ultra accessories

21 (4) Repeat procedure as necessary. Over a period of three or four days of cycling it is frequently possible to restore a significant portion of l

Página 15 - Chapter 2 Basic operation

22 CAUTION: Never perform remote sampling with the PhD Ultra without the sample probe assembly. The sample probe handle contains replaceable filte

Página 16 - 2.2 Operating modes

23 2.4.2.2 Protective “low flow” shut-downs If a protective pump shut-down occurs, the following steps should be taken before the instrument is put

Página 17 - 2.2.3 Technician mode

24 new tube, inserting the new tube into the probe handle, and finally replacing and re-tightening the hex-nut. Note: The sample probe must be check

Página 18 - 2.3 Batteries

25 Chapter 3 Calibration The PhD Ultra multi-gas detector has been designed for easy calibration. A single control, the on/off mode button, is used

Página 19 - 2.3.3 Low battery alarms

26 Using Biosystems brand calibration gas and regularly verifying accuracy ensures that proper sensitivity is maintained for the life of the sensor.

Página 20

27 space entry (29 CFR 1910.146) explicitly requires the use of direct reading, substance specific sensors whenever a particular toxic hazard is know

Página 21 - 2.4 Methods of sampling

28 Since fresh air contains 20.9 percent oxygen, the fresh air “zero” calibration is the only calibration needed for the oxygen sensor in the PhD Ult

Página 22

29 (4) Press the mode button three times within two seconds. This will "wake up" the instrument from normal operation, and put it into th

Página 23 - 2.4.3 Sample probe assembly

3 Table of Contents INTRODUCTION 7 SIGNAL WORDS 7 WARNINGS 8 CHAPTER 1 PHD ULTRA DESCRIPTION 9 1.1 PHD ULTRA CAPABILITIES 9 1.2 METHODS OF SAM

Página 24 - “Smart Sensors”

30 Figure 3.4.2. PhD Ultra calibration setup The regulator will automatically begin to flow calibration gas as soon as it is screwed into the cyl

Página 25 - Chapter 3 Calibration

31 Wait at least 3 minutes after turning the instrument on to allow sensor readings to stabilize fully before initiating any calibration procedures

Página 26

32 Note: Make sure to use the calibration / sample draw adapter supplied with the hand aspirated sample draw assembly. Do not use the battery opera

Página 27

33 Chapter 4 PhD Ultra Advanced Functions 4.1 PhD Ultra advanced features overview PhD Ultra microprocessor circuitry makes a number of advanced fe

Página 28 - 3.4 Auto-calibration

34 4.2.2 Viewing current or restoring the factory default alarm settings PhD Ultra alarm settings are set very conservatively at the factory. (Se

Página 29

35 Alarm latch Security beep, Toxic sensor decimal point, Calibration gas settings User ID number. Each of these options is described in greater

Página 30

36 PhD Ultra alarms are normally self-resetting, which means that the alarms cease as soon as reading drops below the alarm set point. PhD Ultra ala

Página 31

37 to scroll through the setup options until you reach the combustible sensor screen. (2) Press the “+” or “-“ keys to change the combustible from

Página 32

38 chosen in the calibration gas setting option, the instrument will display H2S+ and hydrogen sulfide must be used to verify accuracy. Use of the i

Página 33 - Functions

39 This screen will be followed by a screen showing the first instrument setup choices. Note: Pressing the mode button at any time cancels the

Página 34 - 4.3 Instrument setup

4 2.2.1 Text Only mode 16 2.2.2 Basic mode 17 2.2.3 Technician mode 17 2.2.3.1 PEAK READINGS 17 2.2.3.2 STEL READINGS 17 2.2.3.3 TWA READI

Página 35

40 “manual” PhD Ultra setup and download procedures. Consult the Datalink Reference Manual for complete instructions in the use of Datalink software

Página 36 - 4.3.8 Operating mode

41 To quit and save the new settings press and hold the “CAL” button. Release the “CAL” button when the instrument instructs you to release it. The

Página 37

42 4.5.3.6 Exiting the Datalogging Adjust mode When the PhD Ultra is properly configured, press and hold the CAL button to save the changes. You wi

Página 38 - CAUTION:

43 will not be retained in the location or user list for future use. This ID feature is available for use while the instrument is operated in any o

Página 39 - 4.5 Record keeping

44 If a location has not been assigned, the screen will be blank other than showing “ID=” or Follow the directions above for the user name and ent

Página 40

45 To continue in the passcode setup option, enter the passcode. If this is the first time entering the passcode option or the passcode has been sav

Página 41 - Make sure that any session

46 Chapter 5 Trouble-shooting and repair Repair procedures may only be performed by authorized personnel! 5.1 Changing PhD Ultra sensors The PhD Ul

Página 42

47 reading “Too Low” or “Too High” for One-Button Auto-Calibration fresh air adjustment. Once the new sensor has been fresh-air calibrated using the

Página 43

48 using the procedures discussed in Section 5.2.1. If the instrument still fails to turn on, return to factory for repair. 5.2.2.3 Problem: Sensor

Página 44 - 4.6 Passcode Overview

49 (2) Remove the battery pack. (3) Remove the sensor grill cover by removing the three screws on the front of the instrument to access the sensor

Página 45 - 4.7 Software / Flash Upload

5 3.5.2 Span calibration using keypad buttons 31 CHAPTER 4 PHD ULTRA ADVANCED FUNCTIONS 33 4.1 PHD ULTRA ADVANCED FEATURES OVERVIEW 33 4.2 SETT

Página 46

50 LCD display indicates that the pump is attached and in normal operation.) The sample draw pump includes a unique pressure sensor designed to pro

Página 47 - 5.2 Troubleshooting

51 Figure 5.4.3. Cutaway views of 54-05-A0101 motorized pump showing major component assemblies 5.4.2 Specific problems with motorized pump 5.4.2.

Página 48 - 5.3 Changing the PhD Ultra

52 Appendices Appendix A Toxic gas measurement - Ceilings, TWAs and STELs Many toxic substances are commonly encountered in industry. The presence

Página 49

53 Appendix B How to determine where to set your alarms 1. Oxygen alarms Two oxygen alarm set points have been provided; one for low concentrations

Página 50

54 shows the "relative response curves" of the instrument to several different gases. CALIBRATIONSTANDARD1008060402020406080100CALIBRATION

Página 51

55 responsive to ethanol than to methane, so the readings will be low. Multiplying the instrument reading by the correction factor for ethanol will

Página 52 - Appendices

56 In the following table "NA" indicates no value has been assigned by OSHA. Note: Customers should be aware that OSHA permissible exposu

Página 53

57 Appendix D Suggested Calibration Gases Use of non-standard calibration gas and/or calibration kit components when calibrating the PhD Ultra ca

Página 54

58 Appendix E PhD Ultra Toxic Sensor Cross Sensitivity Data1 The table below lists the cross sensitivity of electrochemical toxic sensors used in B

Página 55

59 Appendix F Calibration Frequency One of the most common questions that we are asked at Biosystems is: “How often should I calibrate my gas detect

Página 56 - Ultra in contaminated air

6 5.2.2.2 PROBLEM: UNIT WILL NOT TURN OFF 47 5.2.2.3 PROBLEM: SENSOR READINGS UNSTABLE IN A KNOWN FRESH AIR ENVIRONMENT 48 5.2.2.4 PROBLEM: &quo

Página 57

60 Appendix G Biosystems Standard Warranty Gas Detection Products General Biosystems LLC (hereafter Biosystems) warrants gas detectors, sensors and a

Página 58 - Appendix E PhD Ultra

7 Introduction The PhD Ultra is a personal, portable, microprocessor controlled gas detector that can monitor up to four atmospheric hazards simultan

Página 59 - Frequency

8 Warnings 1. The PhD Ultra personal, portable gas detector has been designed for the detection of dangerous atmospheric conditions. An alarm cond

Página 60 - General

9 Chapter 1 PhD Ultra Description 1.1 PhD Ultra capabilities The PhD Ultra gas detector can be configured to meet a wide variety of requirements.

Comentários a estes Manuais

Sem comentários